HygroPro
II
User's Manual
51
Appendix A. HygroPro
II
HART
®
Field Device Specification
A.9.4 Command 133 (0x85): Set Primary Variable Range Values
This command is to set the PV range.
Request Data Bytes
Response Data Bytes
Command-Specific Response Codes
Byte
Format
Description
0
Unsigned-8
Upper and Lower Range Values Unit Class
1
Unsigned-8
Upper and Lower Range Values Unit Code
2 ~ 5
Float
Upper Range Value
6 ~ 9
Float
Lower Range Value
Byte
Format
Description
0
Unsigned-8
Upper and Lower Range Values Unit Class
1
Unsigned-8
Upper and Lower Range Values Unit Code
2 ~ 5
Float
Upper Range Value
6 ~ 9
Float
Lower Range Value
Code
Class
Description
0
Success
No Command-Specific Errors
1
Undefined
2
Error
Invalid Selection
3 ~ 4
Undefined
5
Error
Too Few Data Bytes Received
6
Error
Device-Specific Command Error
7
Error
In Write Protect Mode
8
Warning
Set To Nearest Possible Value (Upper or Lower Range Pushed)
9
Error
Lower Range Value Too High
10
Error
Lower Range Value Too Low
11
Error
Upper Range Value Too High
12
Error
Upper Range Value Too Low
13
Error
Upper and Lower Range Values Out Of Limits
14
Warning
Span Too Small
15
Undefined
16
Error
Access Restricted
17
Undefined
18
Error
Invalid Units Code
19 -28
Undefined
29
Error
Invalid Span
30 ~ 31
Undefined
Содержание HydroProII
Страница 1: ...panametrics com 910 331 A November 2022 HygroProII Moisture Transmitter User s Manual...
Страница 2: ......
Страница 4: ...ii no content intended for this page...
Страница 12: ...Preface x HygroProII User s Manual no content intended for this page...
Страница 42: ...Chapter 2 Operation 30 HygroProII User s Manual no content intended for this page...
Страница 52: ...Chapter 4 Specifications 40 HygroProII User s Manual no content intended for this page...
Страница 100: ...Appendix B HygroProII Field Service Update User Manual 88 HygroProII User s Manual no content intended for this page...
Страница 102: ...Warranty 90 HygroProII User s Manual no content intended for this page...
Страница 103: ......