HygroPro
II
User's Manual
49
Appendix A. HygroPro
II
HART
®
Field Device Specification
A.9.2 Command 131 (0x83): Set Loop Current Gain
This command is to trim the zero or low endpoint value to a value of 4 mA.
For example, if the zero value is 4.1 mA, then the correction entered should be -0.1 mA to achieve a result of 4 mA.
Request Data Bytes
Response Data Bytes
Command-Specific Response Codes
32
Error
Busy
33 ~ 127
Undefined
Byte
Format
Description
0
Unsigned-8
Enumeration. 0 = Fixed 20 mA trim, 1 = Live mA
Set for live analog output or fixed 20 mA for trim purposes
1~4
Float
mA correction value
Byte
Format
Description
0
Unsigned-8
Enumeration. 0 = Fixed 20 mA trim, 1 = Live mA
Set for live analog output or fixed 20 mA for trim purposes
1~4
Float
mA correction value
Code
Class
Description
0
Success
No Command-Specific Errors
1-2
Undefined
3
Error
Passed Parameter Too Large
4
Error
Passed Parameter Too Small
5
Error
Too Few Data Bytes Received
6
Error
Device-Specific Command Error
7
Error
In Write Protect Mode
8
Undefined
9
Error
Incorrect Loop Current Mode or Value
10
Undefined
11
Error
Loop Current Not Active
12 ~ 15
Undefined
16
Error
Access Restricted
17-31
Undefined
32
Error
Busy
33 ~ 127
Undefined
Code
Class
Description
Содержание HydroProII
Страница 1: ...panametrics com 910 331 A November 2022 HygroProII Moisture Transmitter User s Manual...
Страница 2: ......
Страница 4: ...ii no content intended for this page...
Страница 12: ...Preface x HygroProII User s Manual no content intended for this page...
Страница 42: ...Chapter 2 Operation 30 HygroProII User s Manual no content intended for this page...
Страница 52: ...Chapter 4 Specifications 40 HygroProII User s Manual no content intended for this page...
Страница 100: ...Appendix B HygroProII Field Service Update User Manual 88 HygroProII User s Manual no content intended for this page...
Страница 102: ...Warranty 90 HygroProII User s Manual no content intended for this page...
Страница 103: ......