HygroPro
II
User's Manual
59
Appendix A. HygroPro
II
HART
®
Field Device Specification
A.9.17 Command 156 (0x9C): Set Pressure Curve
This command sets the two point pressure curve.
Request Data Bytes
Response Data Bytes
Command-Specific Response Codes
A.9.18 Command 157 (0x9D): Read Pressure Curve
This command reads the two point pressure curve.
Request Data Bytes
Response Data Bytes
Byte
Format
Description
0 ~ 3
Float
Low End mV
4 ~ 7
Float
Low End PSIg
8 ~ 11
Float
High End mV
12 ~ 15
Float
High End PSIg
Byte
Format
Description
0 ~ 3
Float
Low End mV
4 ~ 7
Float
Low End PSIg
8 ~ 11
Float
High End mV
12 ~ 15
Float
High End PSIg
Code
Class
Description
0
Success
No Command-Specific Errors
1-4
Undefined
5
Error
Too Few Data Bytes Received
6
Error
Device-Specific Command Error
7
Error
In Write Protect Mode
8-15
Undefined
16
Error
Access Restricted
8-127
Undefined
Byte
Format
Description
None
Byte
Format
Description
0 ~ 3
Float
Low End mV
4 ~ 7
Float
Low End PSIg
Содержание HydroProII
Страница 1: ...panametrics com 910 331 A November 2022 HygroProII Moisture Transmitter User s Manual...
Страница 2: ......
Страница 4: ...ii no content intended for this page...
Страница 12: ...Preface x HygroProII User s Manual no content intended for this page...
Страница 42: ...Chapter 2 Operation 30 HygroProII User s Manual no content intended for this page...
Страница 52: ...Chapter 4 Specifications 40 HygroProII User s Manual no content intended for this page...
Страница 100: ...Appendix B HygroProII Field Service Update User Manual 88 HygroProII User s Manual no content intended for this page...
Страница 102: ...Warranty 90 HygroProII User s Manual no content intended for this page...
Страница 103: ......