Chapter 1. Installation
12
HygroPro
II
User's Manual
Figure 12: Hazardous Area connections with Analog output or HART device
If using serial communications with a PC, an isolated RS232-RS485 converter mounted in the safe area between the
computer and HygroPro
II
must be used. The converter is typically powered by its own standard 24 VDC power supply.
WARNING!
Do NOT power an RS232-RS485 converter from the same intrinsically safe 24 VDC power
supply used to power the HygroPro
II
.