HygroPro
II
User's Manual
61
Appendix A. HygroPro
II
HART
®
Field Device Specification
Response Data Bytes
Command-Specific Response Codes
A.9.21 Command 160 (0xA0): Set Saturation Table Point
Sets a saturation table point by index.
Request Data Bytes
Response Data Bytes
Command-Specific Response Codes
Byte
Format
Description
0
Unsigned-8
Number of Saturation Table Points.
Code
Class
Description
0
Success
No Command-Specific Errors
1-127
Undefined
Byte
Format
Description
0
Unsigned-8
Index of saturation table point.
1 ~ 4
Float
Temperature degC
5 ~ 8
Float
Saturation Constant
Byte
Format
Description
0
Unsigned-8
Index of saturation table point.
1 ~ 4
Float
Temperature degC
5 ~ 8
Float
Saturation Constant
Code
Class
Description
0
Success
No Command-Specific Errors
1-3
Undefined
4
Error
Index Out of Bounds Error.
5
Error
Too Few Data Bytes Received
6
Error
Device-Specific Command Error
7
Error
In Write Protect Mode
8-15
Undefined
16
Error
Access Restricted
8-127
Undefined
Содержание HydroProII
Страница 1: ...panametrics com 910 331 A November 2022 HygroProII Moisture Transmitter User s Manual...
Страница 2: ......
Страница 4: ...ii no content intended for this page...
Страница 12: ...Preface x HygroProII User s Manual no content intended for this page...
Страница 42: ...Chapter 2 Operation 30 HygroProII User s Manual no content intended for this page...
Страница 52: ...Chapter 4 Specifications 40 HygroProII User s Manual no content intended for this page...
Страница 100: ...Appendix B HygroProII Field Service Update User Manual 88 HygroProII User s Manual no content intended for this page...
Страница 102: ...Warranty 90 HygroProII User s Manual no content intended for this page...
Страница 103: ......