Appendix A. HygroPro
II
HART
®
Field Device Specification
54
HygroPro
II
User's Manual
A.9.8 Command 147 (0x93): Read Pressure Constant
This command reads the pressure constant used in place of a live pressure measurement.
First byte is an enumeration: 0 = pressure constant active, 1 = live measurement instead of constant
Request Data Bytes
Response Data Bytes
Command-Specific Response Codes
A.9.9 Command 148 (0x94): Set Temperature Constant
Sets a temperature constant to be used in place of a live measurement.
First byte is an enumeration: 0 = temperature constant active, 1 = live measurement instead of constant
Request Data Bytes
Response Data Bytes
16
Error
Access Restricted
8-127
Undefined
Byte
Format
Description
None
Byte
Format
Description
0
Unsigned-8
0 = use constant 1 = use live
1 ~ 4
Float
Pressure constant used instead of live measurement.
Code
Class
Description
0
Success
No Command-Specific Errors
1-127
Undefined
Byte
Format
Description
0
Unsigned-8
0 = use constant 1 = live
1~4
Float
Temperature constant value.
Byte
Format
Description
0
0 = use constant 1 = live
1~4
Float
Temperature constant value.
Code
Class
Description
Содержание HydroProII
Страница 1: ...panametrics com 910 331 A November 2022 HygroProII Moisture Transmitter User s Manual...
Страница 2: ......
Страница 4: ...ii no content intended for this page...
Страница 12: ...Preface x HygroProII User s Manual no content intended for this page...
Страница 42: ...Chapter 2 Operation 30 HygroProII User s Manual no content intended for this page...
Страница 52: ...Chapter 4 Specifications 40 HygroProII User s Manual no content intended for this page...
Страница 100: ...Appendix B HygroProII Field Service Update User Manual 88 HygroProII User s Manual no content intended for this page...
Страница 102: ...Warranty 90 HygroProII User s Manual no content intended for this page...
Страница 103: ......