Appendix B. HygroPro
II
Field Service Update User Manual
14. If new instrument program firmware was installed the bootloader will return to the main menu
15. If new bootloader firmware was downloaded, then the device will automatically restart and run the new boot
-
loader
a.
NOTE: You will have to enter the password again
B.3.4 Erase NVM Menu
The Erase Non-Volatile Memory Menu is used to erase any settings or calibration data stored in the flash
memory of the device. For the HygroPro
II
Probe this includes all calibrations, calibration date and serial numbers. For
the HygroPro
II
Display this includes all 4-20mA output settings, display settings and serial numbers.
B.3.4.1 To erase non-volatile memory
1.
From the Main Menu press “3” to open the Erase NVM Menu
2. Press “y” to confirm
3. The device will attempt to erase and show a status message
4. IF the device is successful it will return to the Main Menu
a.
If the device is unsuccessful it will return to the Erase NVM Menu
b.
You may press “y” to attempt the erase again
Figure 35: Erase Non-Volatile Memory Menu
Содержание HydroProII
Страница 1: ...panametrics com 910 331 A November 2022 HygroProII Moisture Transmitter User s Manual...
Страница 2: ......
Страница 4: ...ii no content intended for this page...
Страница 12: ...Preface x HygroProII User s Manual no content intended for this page...
Страница 42: ...Chapter 2 Operation 30 HygroProII User s Manual no content intended for this page...
Страница 52: ...Chapter 4 Specifications 40 HygroProII User s Manual no content intended for this page...
Страница 100: ...Appendix B HygroProII Field Service Update User Manual 88 HygroProII User s Manual no content intended for this page...
Страница 102: ...Warranty 90 HygroProII User s Manual no content intended for this page...
Страница 103: ......