Chapter 3. Logging Data
68
DigitalFlow™ DF868 Multipurpose Ultrasonic Liquid Flowmeter Programming Manual (2-Channel)
3.2.3 START DATE Prompt
11.
Press
[F1]
to accept the displayed
Start Date
or press
[F2]
to enter a different start date. To start logging today,
press
[F3]
.
• If
OK
or
TODAY
was selected, proceed to step 12 for a non-circular log or to step 14 for a circular log.
• Otherwise, continue with the next prompt to
Edit
the Start Date.
a.
Enter the desired
Year
and press
[ENT]
. The acceptable range is 0 to 99.
b.
Use the [
, [
and
[F1]-[F4]
keys to select the desired
Month.
c.
Enter the desired
Day
and press
[ENT]
. The acceptable range is 1 to the number of days in the selected
month (28, 29, 30 or 31).
3.2.4 END TIME Prompt
12.
Press
[F1]
to accept the displayed
End Time
or press
[F2]
to enter a different end time. To select a specific time
period for the log to run, press
[F3]
.
•
If
OK
was selected, proceed to the
END DATE
prompt.
•
If
TIMED
was selected, proceed to the
LOG TIME
prompt.
•
If
EDIT
was selected, complete the steps below.
Note:
The log
END TIME
must exceed the log
START TIME
by at least five minutes. Failure to observe this restriction
will result in an error message.
a.
Press
[F1]
for
AM
or press
[F2]
for
PM
. Then, enter the desired
Hour
and press
[ENT]
. The acceptable range is 1 to 12.
b.
Enter the desired
Minutes
and press
[ENT]
. The acceptable range is 0 to 59.
c.
Enter the desired
Seconds
and press
[ENT]
. The acceptable range is 0 to 59.
3.2.5 END DATE Prompt
13.
Press
[F1]
to accept the displayed
End Date
or press
[F2]
to enter a different end date. To end the log today, press
[F3]
.
• If
OK
or
TODAY
was selected, proceed to the
TIME INCREMENT
prompt.
• Otherwise, complete the steps below.
a.
Enter the desired
Year
and press
[ENT]
. The acceptable range is 0 to 99.
b.
Use the [
, [
and
[F1]-[F4]
keys to select the desired
Month
.
c.
Enter the desired
Day
and press
[ENT]
. The acceptable range is 1 to the number of days in the selected
month (28, 29, 30 or 31). The program proceeds to step 15.
Содержание DigitalFlow DF868
Страница 2: ......
Страница 4: ...ii no content intended for this page...
Страница 10: ...Contents viii DigitalFlow DF868 Programming Manual 2 Channel...
Страница 104: ...Appendix A Menu Maps 96 DigitalFlow DF868 Multipurpose Ultrasonic Liquid Flowmeter Programming Manual 2 Channel...
Страница 118: ...Appendix B Data Records 110 DigitalFlow DF868 Multipurpose Ultrasonic Liquid Flowmeter Programming Manual 2 Channel...
Страница 150: ...Appendix E OPC Server 142 DigitalFlow DF868 Multipurpose Ultrasonic Liquid Flowmeter Programming Manual 2 Channel...
Страница 158: ...2 DigitalFlow DF868 Multipurpose Ultrasonic Liquid Flowmeter Programming Manual 2 Channel...
Страница 159: ......