Chapter 3. Logging Data
66
DigitalFlow™ DF868 Multipurpose Ultrasonic Liquid Flowmeter Programming Manual (2-Channel)
3.2 Creating a Standard Log
Use the
STD
submenu to create a new standard log and to select the parameters to log, the log start time and date,
the log end time and date, and the time increment. Also, any log file already stored in memory may be inspected
and/or changed.
Press the appropriate side of the
[SCREEN]
key to activate the desired pane of the display screen, and complete the
following steps:
1.
To access the
Log Menu
, press
[LOG]
.
a.
If you have activated the security feature (see Chapter 1,
Programming Site Data
, page 50 for details), enter the
assigned password and press
[ENT]
.
2.
Press
[F1]
=
STD
.
3.
Use the [
, [
and
[F1]-[F4]
keys to inspect a completed log, change an active log, or enter a new name (up
to 5 characters) and press
[ENT]
to create a new log file.
Note:
At the
NAME
prompt, the option bar will show the names of all completed or active log files that are
currently stored in memory. Recall that a flashing asterisk (*) appears at the far right of the locator bar if the
Model DF868 is currently logging data (see Chapter 3,
Operation
, page 3-4 of the
Startup Guide
).
4.
At the
Log Message
prompt, enter a brief description (up to 21 characters) of the log and press
[ENT]
.
5.
Use the [
, [
and
[F1]-[F4]
keys to select the first parameter to be logged. See Table 35 on page 66 for a list
of the available options.
IMPORTANT:
Only the first parameter logged may be graphically displayed in the
LOG
submenu of the Display Menu.
Keep this in mind when selecting a parameter. See Chapter 2, Displaying Data, for instructions.
Table 35: Measurement Parameter Options
Option Bar Choice
Description
[F1]
=
VEL
Flow Velocity
[F2]
=
VOLUM
Volumetric Flow
[F3]
=
+TOTL
Forward Totalized Volume Flow
[F4]
=
-TOTL
Reverse Totalized Volume Flow
[
+ [F1]
=
MDOT*
Mass Flow
[
+ [F2]
=
+MASS*
Forward Totalized Mass Flow
[
+ [F3]
=
-MASS*
Reverse Totalized Mass Flow
[
+ [F4]
= POWER**
Energy Flow Power
[
+ [
+ [F1]
=
+ENRG**
Forward Energy Flow
[
+ [
+ [F2]
=
-ENRG**
Reverse Energy Flow
[
+ [
+ [F3]
=
DIAG
Diagnostic
* Available only if “Mass Flow (static)?” = YES
** Available only if “Energy Option?” = YES
Содержание DigitalFlow DF868
Страница 2: ......
Страница 4: ...ii no content intended for this page...
Страница 10: ...Contents viii DigitalFlow DF868 Programming Manual 2 Channel...
Страница 104: ...Appendix A Menu Maps 96 DigitalFlow DF868 Multipurpose Ultrasonic Liquid Flowmeter Programming Manual 2 Channel...
Страница 118: ...Appendix B Data Records 110 DigitalFlow DF868 Multipurpose Ultrasonic Liquid Flowmeter Programming Manual 2 Channel...
Страница 150: ...Appendix E OPC Server 142 DigitalFlow DF868 Multipurpose Ultrasonic Liquid Flowmeter Programming Manual 2 Channel...
Страница 158: ...2 DigitalFlow DF868 Multipurpose Ultrasonic Liquid Flowmeter Programming Manual 2 Channel...
Страница 159: ......