Chapter 1. Programming Site Data
38
DigitalFlow™ DF868 Multipurpose Ultrasonic Liquid Flowmeter Programming Manual (2-Channel)
Programming a Frequency Output
This choice produces a frequency pulse that is proportional to the output measurement.
4.
Use the [
and [
and
[F1]-[F4]
keys to specify the desired
Output Measurement
parameter. See Table 21 on
page 36 for a description of the available options.
Note:
The measurement units that appear in these prompts are those selected in the
SYSTM
menu earlier in this
chapter.
5.
Enter a value for the
Base
(low end) of the frequency output range and press
[ENT]
.
6.
Enter a value for the
Full
(high) end of the frequency output range and press
[ENT]
.
7.
Enter a value between 1 and 10,000 for the
Full Scale Frequency
and press
[ENT]
.
8.
Use the
[F1]-[F4]
keys to select the desired
Error Option
, the way the display responds during an error condition.
Available options are shown in Table 23. Then skip to step 12 on the next page.
Table 23: Error Response Options for Frequency Output
Option
Display Response
HOLD
Holds the last “good” reading.
LOW
Forces the outputs to the low setpoint.
HIGH
Forces the outputs to the high setpoint.
OTHER
Allows user to enter Hertz error level.
Содержание DigitalFlow DF868
Страница 2: ......
Страница 4: ...ii no content intended for this page...
Страница 10: ...Contents viii DigitalFlow DF868 Programming Manual 2 Channel...
Страница 104: ...Appendix A Menu Maps 96 DigitalFlow DF868 Multipurpose Ultrasonic Liquid Flowmeter Programming Manual 2 Channel...
Страница 118: ...Appendix B Data Records 110 DigitalFlow DF868 Multipurpose Ultrasonic Liquid Flowmeter Programming Manual 2 Channel...
Страница 150: ...Appendix E OPC Server 142 DigitalFlow DF868 Multipurpose Ultrasonic Liquid Flowmeter Programming Manual 2 Channel...
Страница 158: ...2 DigitalFlow DF868 Multipurpose Ultrasonic Liquid Flowmeter Programming Manual 2 Channel...
Страница 159: ......