ORTEC
®
Detective
®
X Handheld Radioisotope Identifier
1107231N / 082020
6
1.5 TURNING THE COOLER ON
1.
In the lower right corner, tap the
Settings
button to open the
Settings
page (Figure 12). Then tap
Turn
Cooler ON
. Note that this is an on/off toggle, and the button label changes with the cooler state.
2.
At this point, the “
The Cooler is turned off!
”
error message will no longer be displayed, and the
instrument status will say
Status: Cooling Phase 1
3.
The instrument may be temporarily disconnected from external power and moved to a different location
if desired.
4.
Cooling will continue through four phases before the status message changes to
Status: READY
.
There is no need to continue displaying the Settings page during cooldown, although the user may wish to
tap the
Status
button to display system temperatures and track cooldown progress.
Figure 12. Turn Cooler On
NOTE
The default factory settings will be enabled to automatically start the ‘traditional’ Detective
X user interface upon system power up. This behavior can be modified to start other
onboard applications (i.e., Sleuth, Mobile MCB Server, or Launcher). If the instrument
settings have been modified to start an alternate program, please refer to Chapters 3
(Launcher), 5 (Sleuth), and 6 (Mobile MCB Server).
NOTE
Once turned on, the cooler stays on, even if the Power button is turned off. To completely
shut down the instrument, press the power button on the front of the instrument. A
message will be displayed asking to turn off the cooler. See Section 2.9 for more
information.
Содержание Detective X
Страница 12: ...ORTEC Detective X Handheld Radioisotope Identifier 1107231N 082020 xii THIS PAGE INTENTIONALLY LEFT BLANK...
Страница 32: ...ORTEC Detective X Handheld Radioisotope Identifier 1107231N 082020 20 THIS PAGE INTENTIONALLY LEFT BLANK...
Страница 34: ...ORTEC Detective X Handheld Radioisotope Identifier 1107231N 082020 22 THIS PAGE INTENTIONALLY LEFT BLANK...
Страница 86: ...ORTEC Detective X Handheld Radioisotope Identifier 1107231N 082020 74 THIS PAGE INTENTIONALLY LEFT BLANK...
Страница 116: ...ORTEC Detective X Handheld Radioisotope Identifier 1107231N 082020 104 THIS PAGE INTENTIONALLY LEFT BLANK...
Страница 120: ...ORTEC Detective X Handheld Radioisotope Identifier 1107231N 082020 108 THIS PAGE INTENTIONALLY LEFT BLANK...
Страница 130: ...ORTEC Detective X Handheld Radioisotope Identifier 1107231N 082020 118 THIS PAGE INTENTIONALLY LEFT BLANK...