ORTEC
®
Detective
®
X Handheld Radioisotope Identifier
1107231N / 082020
50
Figure 62. Enable Collection of Multiple Spectra in a Single File
Figure 63. Set Up Parameters for Multiple Spectrum ID Mode
4.7.4.1
Setting Up and Performing Multiple Measurements
The Multi ID Mode page allows the user to define and collect a series of spectra, one at a time, until the
desired data set has been collected. The software then saves all of the spectra and metadata in a single
N42.42 file.
To set up and perform this type of measurement:
1.
Select the desired category of spectrum from the
Next spectrum type
field. The options are
Blank
,
Known
, and
Unknown
. A typical measurement series might consist of one Blank, one Known, and one or
more Unknowns. However, there are no restrictions on what type or how many of each type of spectra
go into the Multi ID Mode measurement file.
2.
Assign a name to this spectrum in the
Next spectrum label
field. This is a free-form field that allows the
entry of information about the spectrum to be collected. The field is initially populated with the text used
Содержание Detective X
Страница 12: ...ORTEC Detective X Handheld Radioisotope Identifier 1107231N 082020 xii THIS PAGE INTENTIONALLY LEFT BLANK...
Страница 32: ...ORTEC Detective X Handheld Radioisotope Identifier 1107231N 082020 20 THIS PAGE INTENTIONALLY LEFT BLANK...
Страница 34: ...ORTEC Detective X Handheld Radioisotope Identifier 1107231N 082020 22 THIS PAGE INTENTIONALLY LEFT BLANK...
Страница 86: ...ORTEC Detective X Handheld Radioisotope Identifier 1107231N 082020 74 THIS PAGE INTENTIONALLY LEFT BLANK...
Страница 116: ...ORTEC Detective X Handheld Radioisotope Identifier 1107231N 082020 104 THIS PAGE INTENTIONALLY LEFT BLANK...
Страница 120: ...ORTEC Detective X Handheld Radioisotope Identifier 1107231N 082020 108 THIS PAGE INTENTIONALLY LEFT BLANK...