ORTEC
®
Detective
®
X Handheld Radioisotope Identifier
1107231N / 082020
68
1.
Place the source in front of the instrument to obtain a gamma count rate of approximately 500 cps.
2.
Tap
Settings
to open the Settings screen, then tap
Calibrate
to display the screen shown in Figure 86.
3.
Tap
Start Calibration Now
. The stabilizer will turn off and the current stabilizer adjustment (if any) will
be reset to zero. The acquisition will begin, and the spectrum will be displayed on the lower left of the
screen.
4.
When enough counts have been accumulated to accurately determine the centroid of the Cs-137 peak,
the system will report the degree of adjustment necessary. Tap
Use Improved Calibration
to complete
the process.
Figure 86. Calibrate Screen
The stabilizer will turn back on after calibration is complete.
The
Restore Default
button reinstalls the factory calibration, which is useful for diagnostic purposes;
however, the factory calibration is typically not optimum for an instrument once it leaves the factory.
4.15.2.1 Using Alternate Calibration Sources
Other sources can be used to energy calibrate the system as well. In fact, environmental K-40 can be used to
calibrate the instrument. This is useful when no Cs-137 check source is available. Calibrating with K-40 can
take a long time in some environments (note that seawater contains essentially no K-40), therefore, the user
may wish to purchase an inexpensive potassium chloride “salt substitute” from a grocery. From the Advanced
Setup screen, tap
Cal Settings
to open the screen shown in Figure 87. Select the desired source from the
Calibration sources
list.
4.15.2.2 Require New Background After Calibration
Although not necessary for the onboard nuclide identification algorithm, it may be desired to force collection
of a new background spectrum each time a new energy calibration is performed. Tap the
Require background
NOTE
If the
Back
button is tapped before tapping
Use Improved Calibration
, the old calibration
will be retained and the new one will be discarded.
Содержание Detective X
Страница 12: ...ORTEC Detective X Handheld Radioisotope Identifier 1107231N 082020 xii THIS PAGE INTENTIONALLY LEFT BLANK...
Страница 32: ...ORTEC Detective X Handheld Radioisotope Identifier 1107231N 082020 20 THIS PAGE INTENTIONALLY LEFT BLANK...
Страница 34: ...ORTEC Detective X Handheld Radioisotope Identifier 1107231N 082020 22 THIS PAGE INTENTIONALLY LEFT BLANK...
Страница 86: ...ORTEC Detective X Handheld Radioisotope Identifier 1107231N 082020 74 THIS PAGE INTENTIONALLY LEFT BLANK...
Страница 116: ...ORTEC Detective X Handheld Radioisotope Identifier 1107231N 082020 104 THIS PAGE INTENTIONALLY LEFT BLANK...
Страница 120: ...ORTEC Detective X Handheld Radioisotope Identifier 1107231N 082020 108 THIS PAGE INTENTIONALLY LEFT BLANK...
Страница 130: ...ORTEC Detective X Handheld Radioisotope Identifier 1107231N 082020 118 THIS PAGE INTENTIONALLY LEFT BLANK...