1107231N / 082020 APPENDIX B. SPECIFICATIONS AND CARE
107
B.5
MOBILE MCB SERVER
The Mobile MCB Server software application enables the Detective X to communicate wired or wirelessly
with ORTEC software applications such as MAESTRO, GammaVision, and Detective® Remote.
Users can develop their own applications through the use of the optional the ORTEC Programmer’s Toolkit
(P/N A11-BW).
B.6
CLEANING INSTRUCTIONS
B.6.1 Cleaning the Exterior
Disconnect the instrument from external power, turn off the cooler, and power off the computer
interface.
Attach the caps on all power and external counter ports. Ensure the communication port cover is in
place and fastened with the captive screws. Finally, make certain the battery covers are in place and
fastened with their captive screws.
Failure to secure all covers will allow water and/or dust to enter
the instrument enclosure!
Remove loose dust on the outside of the instrument with a lint-free cloth.
Remove remaining dirt with a lint-free cloth dampened in a mild detergent and water solution.
Do
not use abrasive cleaners.
Dry the instrument with a towel and turn on the computer and cooler.
Allow the instrument to dry completely
before reconnecting it to the external power supply.
B.6.2 Cleaning the Fan Compartment
The external cooling fan is in a sealed chamber to prevent water ingress through this area, and the fan itself
is sealed to prevent damage from water. If the Detective X has been used in salt spray or extremely dusty
conditions, follow these instructions to remove the residue by pouring a small amount of water through the
fan chamber:
Disconnect the instrument from external power, turn off the cooler, and power off the computer
interface.
POUR small to moderate amounts of water through the fan chamber.
Do not use pressurized water,
as with a hose; and DO NOT IMMERSE!
Allow the instrument to dry completely
before turning on the computer and cooler, then reconnect
it to the external power supply.
B.7
REPLACING THE INTERNAL BATTERIES
The typical service life of the Detective X’s Li-ion batteries is 3–5 years. When the internal batteries no longer
maintain a charge for a significant period, they must be replaced. Replacements are available from ORTEC as
part number DETECTIVE-X-ACC-BAT.
It is strongly recommended to replace the batteries in pairs.
After replacing the batteries (unless they have been pre-charged), the Detective X must be connected to
external power for 3–4 hours to fully charge the new batteries.
Return the old batteries to ORTEC for recycling.
For more information, contact an ORTEC representative or ORTEC’s Global Service Center.
Содержание Detective X
Страница 12: ...ORTEC Detective X Handheld Radioisotope Identifier 1107231N 082020 xii THIS PAGE INTENTIONALLY LEFT BLANK...
Страница 32: ...ORTEC Detective X Handheld Radioisotope Identifier 1107231N 082020 20 THIS PAGE INTENTIONALLY LEFT BLANK...
Страница 34: ...ORTEC Detective X Handheld Radioisotope Identifier 1107231N 082020 22 THIS PAGE INTENTIONALLY LEFT BLANK...
Страница 86: ...ORTEC Detective X Handheld Radioisotope Identifier 1107231N 082020 74 THIS PAGE INTENTIONALLY LEFT BLANK...
Страница 116: ...ORTEC Detective X Handheld Radioisotope Identifier 1107231N 082020 104 THIS PAGE INTENTIONALLY LEFT BLANK...
Страница 120: ...ORTEC Detective X Handheld Radioisotope Identifier 1107231N 082020 108 THIS PAGE INTENTIONALLY LEFT BLANK...
Страница 130: ...ORTEC Detective X Handheld Radioisotope Identifier 1107231N 082020 118 THIS PAGE INTENTIONALLY LEFT BLANK...