5-8
6.
Ref. CH
Set the reference channel (See the explanation).
Confirming operation
To confirm the new settings, press the DISP/ENTER key.
To cancel the new settings, press the ESC key. A window appears for you to confirm the
cancellation. Select [Yes] using the arrow keys and press the DISP/ENTER key.
Explanation
The following table shows the mode, type, range, and measurable range.
Mode
Type
Range
Measurable Range
Delta
Volt
20 mV
–20.00 to 20.00 mV
60 mV
–60.00 to 60.00 mV
200 mV
–200.0 to 200.0 mV
2 V
–2.000 to 2.000 V
6 V
–6.000 to 6.000 V
20 V
–20.00 to 20.00 V
Delta
TC
R
–1760.0 to 1760.0
°
C
S
–1760.0 to 1760.0
°
C
B
–1820.0 to 1820.0
°
C
K
–1570.0 to 1570.0
°
C
E
–1000.0 to 1000.0
°
C
J
–1300.0 to 1300.0
°
C
T
–600.0 to 600.0
°
C
N
–1300.0 to 1300.0
°
C
W
–2315.0 to 2315.0
°
C
L
–1100.0 to 1100.0
°
C
U
–600.0 to 600.0
°
C
Delta
RTD
Pt100
–800.0 to 800.0
°
C
JPt100
–750.0 to 750.0
°
C
CU1 to 6 (CU10)
–500.0 to 500.0
°
C (option)
CU25
–500.0 to 500.0
°
C (option)
Delta
DI
Level
–1 to 1
Contact
–1 to 1
The relationship with the reference channel
Even if the input type or the measurement range of the difference computation channel
and the reference channel is not the same, the difference computation is performed
according to the following rules.
• When the decimal position between the reference channel and the difference
computation channel is different, the measured value of the reference channel is
adjusted to the decimal position of the measured value of the difference computation
channel to make the computation.
Example : When the measured value of the difference computation channel is 10.00
and the measured value of the reference channel is 100.0, the
computation result becomes 10.00 – 100.0 = –90.00.
• When the units for the reference channel and the difference computation channel are
different, the measured value is not adjusted.
Example : When the measured value of the difference computation channel is 10.00 V
and the measured value of the reference channel is 5.00 mV, the
computation result becomes 10.00 V – 5.00 mV = 5.00 V.
• When the reference channel is set to [Scale] or [Sqrt], the computation uses the
scaled values.
5.4 Difference Computation (Delta) Setting.
Содержание RD-MV100 Series
Страница 40: ...1 22 This page intentionally left blank...
Страница 41: ...1 23 Overview of Functions 1 This page intentionally left blank...
Страница 139: ...7 5 Setting and Operating the Display 7 This page intentionally left blank...
Страница 140: ...7 6 This page intentionally left blank...
Страница 215: ...10 15 Other Functions 10 This page intentionally left blank...
Страница 216: ...10 16 This page intentionally left blank...
Страница 217: ...10 17 Other Functions 10 This page intentionally left blank...
Страница 218: ...10 18 This page intentionally left blank...
Страница 219: ...10 19 Other Functions 10 This page intentionally left blank...