Appendix
83
VISERA RHINO-LARYNGO VIDEOSCOPE ENF-V2
Portable RF communications equipment (including
peripherals such as antenna cables and external antennas)
should be used no closer than 30 cm (12 inches) to any part
of the video system center, including cables specified by
Olympus. Otherwise, degradation of the performance of this
equipment could result.
Guidance and manufacturer’s declaration — Cables used
for EMC compliance testing
Refer to the instruction manuals for each piece of equipment.
Test frequency
[MHz]
Band [MHz]
Modulation
1
Maximum power
[W]
Immunity test level
[V/m]
385
380 – 390
Pulse modulation
1
18 Hz
1.8
27
450
430 – 470
FM
5 kHz
deviation 1 kHz sine
2
28
710
704 – 787
Pulse modulation
1
217 Hz
0.2
9
745
780
810
800 – 960
Pulse modulation
1
18 Hz
2
28
870
930
1720
1700 – 1990
Pulse modulation
1
217 Hz
2
28
1845
1970
2450
2400 – 2570
Pulse modulation
1
217 Hz
2
28
5240
5100 – 5800
Pulse modulation
1
217 Hz
0.2
9
5500
5785
1
The carrier shall be modulated using a 50% duty cycle square wave signal.