Chapter 7 Cleaning, Disinfection and Sterilization Procedures for Endoscopic Instruments
63
VISERA RHINO-LARYNGO VIDEOSCOPE ENF-V2
3.
Soak the endoscope for minimum 1 hour at the temperature recommended
by the detergent manufacturer. For information about the concentration and
immersion time, refer to the instructions given by the detergent
manufacturer. Do not exceed the manufacturer’s maximum recommended
levels.
4.
Remove the endoscope from detergent solution.
5.
After soaking the endoscope, manually clean it following the procedures
described in this Section, then disinfect or sterilize the endoscope following
the procedure described in this Section 7.7, “Disinfection” on page 63 or
Section 7.10, “Sterilization” on page 67.
7.7
Disinfection
All disinfection steps should be performed with the
endoscope completely immersed. Otherwise, disinfectant
solution may not adequately contact all surfaces of the
equipment. As a result, the effectiveness of disinfection may
be reduced.
• For information about the concentration, the temperature and
immersion time refer to the instructions given by the
disinfectant solution’s manufacturer. Do not exceed the
manufacturer’s maximum recommended levels.
• Regarding the disinfectant solution, refer to Section 6.4,
“Disinfectant solution” on page 40.
After manual cleaning, disinfect the endoscope according to the procedure
described below.
Equipment needed
Prepare the following equipment and wear appropriate personal protective
equipment.
• Personal protective equipment
• Clean, lint-free cloth
• Large basins
• Disinfectant solution