Chapter 3 Preparation and Inspection
VISERA RHINO-LARYNGO VIDEOSCOPE ENF-V2
4.
\
Insert the video connector with its UP mark facing upwards, into the socket
of the video system center until the lock lever of the video system center
clicks into position, while holding the video system center stationary with
one hand so that it does not move (see Figure 3.4).
Figure 3.4
5.
Confirm that the video connector is securely attached by pulling it gently.
3.4
Inspection of the endoscopic system
Inspection of the endoscopic image
• Do not stare directly at the distal end of the endoscope while
the examination light is ON. Otherwise, eye injury may result.
• Generally reprocessing agents
1
are aggressive and might
effect the distal end of the endoscope. When the effect
progresses, fogging of the inner lenses might appear, which
result in a foggy video image. Any further use of the
endoscope under this condition might lead to lens to come
off. Before inserting the endoscope to a patient, make sure
that there is no fogging by checking the video image, and by
wiping the outer surface of the lens at the distal end of the
endoscope with a soft cloth. In case fogging is found to be
apparent, please stop using the scope immediately.
1.
Before inspection, wipe the objective lens with a clean, lint-free cloth
moistened with 70% ethyl alcohol.
1 Contact Olympus for the names of specific brands that have
been tested for compatibility with the endoscope.