Chapter 3 Preparation and Inspection
21
VISERA RHINO-LARYNGO VIDEOSCOPE ENF-V2
3.2
Preparation and inspection of the endoscope
Clean and disinfect or sterilize the endoscope as described in Chapter 5,
“Reprocessing: General Policy” through Chapter 7, “Cleaning, Disinfection and
Sterilization Procedures for Endoscopic Instruments”.
Inspection of the endoscope
1.
Inspect the control section, video connector and light guide connector
section for excessive scratching, deformation or other irregularities.
2.
Inspect the boot and the insertion tube near the boot for bends, twists or
other irregularities.
3.
Inspect the external surface of the entire insertion tube for dents, bulges,
swelling, peeling or other irregularities.
4.
Holding the insertion tube gently with one hand, carefully run your fingertips
over the entire length of the insertion tube in both directions. Inspect for any
protruding objects or other irregularities. Also confirm that the insertion tube
is not abnormally rigid (see Figure 3.2).