Chapter 3 Preparation and Inspection
23
VISERA RHINO-LARYNGO VIDEOSCOPE ENF-V2
Inspection for smooth operation
1.
Turn the UP/DOWN angulation control lever slowly in each direction until it
stops. Confirm that the bending section angulates smoothly and correctly
and that maximum angulation can be achieved.
2.
Turn the UP/DOWN angulation control lever slowly to its straight (neutral)
position. Confirm that the bending section returns smoothly to an
approximately straight condition.
3.3
Preparation, inspection and connection of
ancillary equipment
Preparation and inspection of ancillary equipment
Prepare and inspect the light source, video system center and display monitor as
described in their respective instruction manuals.
Connection of the endoscope and ancillary equipment
If the video connector and video system center are not
connected properly, the endoscopic image may flicker or may
not be displayed.
Before inserting the video connector, confirm that the
electrical contacts are not wet. If the electrical contacts are
wet, thoroughly wipe and dry the electrical contacts.
Connect the endoscope’s light guide connector to the light
source, then connect the video connector to the video system
center. Otherwise, the video cable may twist excessively.
1.
Confirm that the light source is OFF.
2.
Hold the light source stationary with one hand and insert the light guide
connector completely into the output socket of the light source.
3.
Confirm that the video system center is OFF.