74
5.4 Inspection of IR playback function
MAJ-2308 INSTRUCTION MANUAL
Ch.5
Confirm that IR playback function works properly.
NOTE
The IR playback function can be used by setting the video recorder to auto
playback mode.
For details on the IR playback function, contact Olympus.
Inspecting the cable connection
Display of WLI image (still image)
Confirm that the WLI image (still image) is displayed on the sub-screen of the observation
monitor during IR imaging.
5.4
Inspection of IR playback function
Confirm that the cables connected to the surgical microscope, video recorder and
observation monitor are properly connected according to Section 3.4, “Connection of
video recorder”.
1
Set the filter selector knob to “IR”.
Figure 5.7
NOTE
When operating the filter selector knob immediately after the operation of the
microscope section, the WLI image (still image) may be blurred.
Filter selector knob