3.2 Connection of Infrared and Blue Light Filter Unit
21
MAJ-2308 INSTRUCTION MANUAL
Ch.3
CAUTION
• Be sure to confirm that the caution lines (yellow) of the lock screws (× 2) are
invisible after the lock screws have been tightened. If any of the caution lines
(yellow) are visible, the filter unit is not connected perfectly. If it is used in such a
condition, the filter unit may drop and it may be damaged.
• If the provided Allen screwdriver (1.5 mm width across flats) is twisted during
tightening of the lock screws (× 2), do not apply force any more. Otherwise, the
equipment and/or the provided Allen screwdriver may be damaged.
6
Slide the filter unit all the way in the direction of the arrow.
At this time, confirm that the caution lines (yellow) of the lock screws are visible.
Figure 3.5
CAUTION
Do not slide the filter unit by force while the lock screws are not completely free. If
the filter unit cannot be slid due to the drag of a screw(s), pull the screw(s) so that
both screws are completely free. Otherwise, equipment damage may result.
7
Using the provided Allen screwdriver (1.5 mm width across flats), tighten the lock
screws (× 2) of the filter unit completely. Confirm that the caution lines (yellow) of the
lock screws (× 2) are invisible.
Figure 3.6
Filter unit
Lock
screws
Caution lines (Yellow)
Allen screwdriver
(1.5 mm width across flats)
Lock screws
Filter unit
Caution line (Yellow)
Содержание MAJ-2308
Страница 2: ......
Страница 16: ...12 1 4 Recorder Remote Cable 3P MAJ 2310 sold separately MAJ 2308 INSTRUCTION MANUAL Ch 1...
Страница 20: ...16 2 3 Infrared Imaging Unit Blue Light Imaging Unit MAJ 2308 INSTRUCTION MANUAL Ch 2...
Страница 72: ...68 4 4 Resetting the current user presets MAJ 2308 INSTRUCTION MANUAL Ch 4...