3.1 Precautions for installation and connection
17
MAJ-2308 INSTRUCTION MANUAL
Ch.3
Chapter 3
Installation and Connection
Prepare this instrument and the ancillary equipment according to the purpose of use as shown in
“
System chart” on page 105.
Also, read the instruction manual for each piece of ancillary equipment and connect this instrument
and the ancillary equipment in the order given in this chapter.
WARNING
Before use, be sure to read this chapter and prepare the ancillary equipment
optimally. Otherwise, patient or operator injury may result as well as malfunctions of
the equipment.
CAUTION
• Be sure to turn the connected surgical microscope and the ancillary equipment
OFF before proceeding to the connection work. Otherwise, equipment failure or
malfunction may result.
• Hold this instrument and other ancillary equipment combined with it securely so as
not to drop them. Otherwise, equipment damage may result.
• Do not install this instrument in the proximity of a source of strong electromagnetic
waves (microwave therapy equipment, shortwave therapy equipment, MRI, radio,
cellphone, etc.). Otherwise, this instrument may fail.
• Install and connect this instrument on a level surface from which the surgical
microscope will not move spontaneously and lock the brake by pushing the brake
pedal. If the base moves spontaneously during work, personnel injury may result.
3.1
Precautions for installation and connection
Содержание MAJ-2308
Страница 2: ......
Страница 16: ...12 1 4 Recorder Remote Cable 3P MAJ 2310 sold separately MAJ 2308 INSTRUCTION MANUAL Ch 1...
Страница 20: ...16 2 3 Infrared Imaging Unit Blue Light Imaging Unit MAJ 2308 INSTRUCTION MANUAL Ch 2...
Страница 72: ...68 4 4 Resetting the current user presets MAJ 2308 INSTRUCTION MANUAL Ch 4...