8.2 Troubleshooting guide
103
MAJ-2308 INSTRUCTION MANUAL
Ch.8
E229
Light source
error
A wrong imaging unit is
connected to the BL imaging unit
terminal of the LED light source.
Connect the BL imaging unit to the LED light
source as described in Section 3.3, “Connection
of Infrared Imaging Unit (MAJ-2304)/ Blue Light
Imaging Unit (MAJ-2305)”.
Irregularity in the BL imaging unit
or LED light source.
Connect the BL imaging unit to the LED light
source as described in Section 3.3, “Connection
of Infrared Imaging Unit (MAJ-2304)/ Blue Light
Imaging Unit (MAJ-2305)”.
If the problem still persists, turn OFF the surgical
microscope and LED light source after use,
unplug the power supply plug of the surgical
microscope from the hospital-grade power outlet,
disconnect the other end of the power cord from
the power inlet receptacle, and contact Olympus.
E230
Light source
not connected
Irregularity in communication
between the surgical microscope
and LED light source.
Follow the instructions in Chapter 10,
“Troubleshooting” in the instruction manuals for
the surgical microscope and Chapter 7,
“Troubleshooting” in the instruction manuals for
LED light source.
E231
Light source
error
The connected LED light source
is incompatible with the surgical
microscope in use.
Follow the instructions in Chapter 10,
“Troubleshooting” in the instruction manuals for
the surgical microscope.
E232
Microscope
error
The IR imaging unit is connected
to an extension unit terminal of
the surgical microscope instead of
the IR imaging unit terminal of the
LED light source.
Connect the IR imaging unit to the IR imaging
unit terminal of the LED light source.
The BL imaging unit is connected
to an extension unit terminal of
the surgical microscope instead of
the BL imaging unit terminal of the
LED light source.
Connect the BL imaging unit to the BL imaging
unit terminal of the LED light source.
E237
Observation
change failed
Failure in image mode switching.
Turn OFF the surgical microscope and LED light
source, unplug the power supply plug of the
surgical microscope from the hospital-grade
power outlet, disconnect the other end of the
power cord from the power inlet receptacle, and
contact Olympus.
Code
Error
message
Possible cause
Solution