50
4.2 User presets
MAJ-2308 INSTRUCTION MANUAL
Ch.4
CAUTION
If the “BL image adjustment” button is not displayed, turn the surgical microscope
OFF, be sure to set the power switch of the LED light source to “
(ON)” and turn
the surgical microscope ON again. If the “BL image adjustment” button is still not
displayed, check the connections by referring to Section 3.3, “Connection of
Infrared Imaging Unit (MAJ-2304)/ Blue Light Imaging Unit (MAJ-2305)”. The
surgical microscope has not been turned ON correctly or the BL imaging unit is not
connected properly.
3
Tap your “User ID” button.
Figure 4.45
4
Confirm that the selected user ID is shown in
the user ID window.
Figure 4.46
5
Tap the “BL image adjustment” button. The
setting items are displayed below the “BL
image adjustment” button.
Figure 4.47
User ID
User ID window
BL image adjustment
Содержание MAJ-2308
Страница 2: ......
Страница 16: ...12 1 4 Recorder Remote Cable 3P MAJ 2310 sold separately MAJ 2308 INSTRUCTION MANUAL Ch 1...
Страница 20: ...16 2 3 Infrared Imaging Unit Blue Light Imaging Unit MAJ 2308 INSTRUCTION MANUAL Ch 2...
Страница 72: ...68 4 4 Resetting the current user presets MAJ 2308 INSTRUCTION MANUAL Ch 4...