Test frequency
[MHz]
Band [MHz]
Modulation
*1
*1 The carrier shall be modulated using a 50% duty cycle square wave signal.
Maximum power
(W)
IMMUNITY TEST
LEVEL [V/m]
385
380 – 390
Pulse modulation
*1
18 Hz
1.8
27
450
430 – 470
FM
*2
5 kHz
deviation 1 kHz sine
*2 As an alternative to FM modulation, 50% pulse modulation at 18 Hz may be used because while it does not
represent actual modulation, it would be worst case.
2
28
710
704 – 787
Pulse modulation
*1
217 Hz
0.2
9
745
780
810
800 – 960
Pulse modulation
*1
18 Hz
2
28
870
930
1720
1700 – 1990
Pulse modulation
*1
217 Hz
2
28
1845
1970
2450
2400 – 2570
Pulse modulation
*1
217 Hz
2
28
5240
5100 – 5800
Pulse modulation
*1
217 Hz
0.2
9
5500