Test Programs
Maintenance Manual
3-24
Page
Test No./
Test Name
Function Outline
•
Transfer speed <Speed> SP1 to SP5
A-arm horizontal speed
<A-arm linier speed> ..................... High speed <Fast> / Medium speed <Middle> / Low speed <Slow>
A-arm rotation speed
<A-arm rotation speed> ................. High speed <Fast> / Medium speed <Middle> / Low speed <Slow>
A-arm vertical speed
<A-arm up down speed> ............... High speed <Fast> / Medium speed <Middle> / Low speed <Slow>
L-arm horizontal speed
<L-arm rotation speed>.................. High speed <Fast> / Medium speed <Middle> / Low speed <Slow>
Macro table rotation speed <Macro table rotation speed> ........ High speed <Fast> / Medium speed <Middle> / Low speed <Slow>
L-arm vertical speed
<L-arm up down speed> ................ High speed <Fast> / Medium speed <Middle> / Low speed <Slow>
Macro table vertical speed <Macro table up down speed>....... High speed <Fast> / Medium speed <Middle> / Low speed <Slow>
No. 25
Contact
Centering
open/close
movement
check
<Contact centering open/close check>
The movement can be checked only if the optional AL120-CC is mounted.
Open and Close are switched between each time the [Start] button is pressed.
No. 26
Software
version
and
error log
display
<Error log Software version>
This program can be used to check the software version and the history of error occurrences.
The software version or error log displayed can be cleared and the error log date can be set by
manipulating the menu buttons on the liquid crystal panel.
[M1]: Clear <Clear>
[M2]: Set date <Date>
[M3]: Go to the previous page <PageUp>
[M4]: Go to the next page <PageDn>
Move the cursor using the 4-way buttons [UP], [DOWN], [RIGHT] and [LEFT].
Press the [Start] button.
The software version and error log are displayed on the liquid crystal panel.
[M1]: Press the <Clear> button and then the [Start] button to delete the error log
Software version <Ver.>: 01.00
Error code display:
Error code <Code>: Refer to the Operation Manual for error code details.
Sequence in which the error occurred <Seq>: Number of the sequence in which the error occurred
Step in which the error occurred <Step>: Number of the step in which the error occurred
Date and time of error occurrence <Date><Time>
•
Press the[Quit] button to quit this program.
•
Up to ten errors are recorded in the error log.
•
Data is cleared after the loader has remained off for continuous two weeks.
•
If the date has been cleared, use [M2]: <Date> to set the date.
•
To output the error log in text format, use the specially designed maintenance software for use with
a PC
Содержание AL120-6Series
Страница 2: ...This page intentionally left blank ...
Страница 9: ...Operation Maintenance Manual Introduction Safety Precautions i 7 Page AL120 86 Series Unit mm ...
Страница 16: ...Introduction Conformity Standards Operation Maintenance Manual i 14 Page ...
Страница 17: ...Operation Maintenance Manual Introduction Conformity Standards i 15 Page ...
Страница 18: ...Introduction Conformity Standards Operation Maintenance Manual i 16 Page This page intentionally left blank ...
Страница 20: ...Maintenance Manual Table of Contents C 2 This page intentionally left blank Page ...
Страница 56: ...Test Programs Maintenance Manual 3 26 Page This page intentionally left blank ...
Страница 64: ...Replacement of Consumables Maintenance Manual 4 8 Page This page intentionally left blank ...
Страница 65: ...This page intentionally left blank ...