Maintenance Manual
Test Programs
3-23
Page
Test No./
Test Name
Function Outline
No. 24
Wafer
type
setting
<Inspection wafer classification setting>
This program can be used to register wafer types, and set and change the size, thickness and transfer
speed for each wafer type.
Press the [Start] button.
The registered wafer types are displayed on the liquid crystal panel.
Specify the item to modify or add, and press the 4-way button [RIGHT].
•
Modify: Move the <
×
> to the item to modify by using the 4-way button.
•
Add: Move the <
×
> to the item to add by using the 4-way button.
Move to the setting item by using the 4-way button, and change the value by using the up and down
buttons.
Move to the comment item by using the 4-way button, and changes into the comment input
screen in the [M3] button <Edit>.
Settable items
•
Wafer size <Size>
:200/150/125/100/---
Normally, the wafer size can be set to any size appropriate for your wafer loading operation.
Standard settings: 150 for AL120-6 series, 200 for AL120-8 series
•
When chose ---, the set point is deleted, and it is not displayed to a wafer type screen at the time of
the inspection.
•
Wafer thickness <Thick>
・
Wafer size <Size>150/125/100mm
:
675-400/400-150
・
Wafer size <Size>200mm
:
725-400/400-180/180-90
* Wafer size and thickness settings may be unavailable, depending on the loader specifications.
•
Transfer speed <Speed>
:Fast/Middle/Slow/SP1
–
SP5
SP1 to SP5 can be set to desired speeds for each axis.
Optional speed is limited as follows by wafer thickness <Thick>.
725-400 ...................................:Fast/Middle/Slow/SP1~SP5 Normal setting<Fast>
400-180 ...................................:Middle/Slow/SP1~SP5 Normal setting <Middle>
180-90 ......................................:Slow/SP1~SP5 Normal setting <Slow>
•
Comment <Comment> ----------- :
A-Z, a-z, 1 to 0, space
[M1]:Decision<OK> [M2]:Cancel<Cancel> [M3]:Back space<BkSpace> [M4]:Exchange<Change>
Capital letter input..... [No.1~No.26]:Input of A to Z
Small letter input ........ [No.1~No.26]:Input of a to z
Number input ................ [No.1~No.9]:Input of 1 to 9, [No.10]:Input of 0, [No.11]:Input of -, [No.12~
No.26]:Input of space
Capital letter input, small letter input, number input are replaced whenever they push [M4]:Exchange
<Change>.
Up to 16 characters can be entered. An input is possible even if it uses PC and the specially
designed maintenance software.
4-way button [UP], [DOWN]: Change the set value of each item
4-way button [LEFT], [RIGHT]: Change the setting item
[M1]: Register <Save>
[M2]: Cancel <Cancel>
[M3]: Detailed settings
When the transfer speed <Speed> is set to SP1 to SP5, press [M3] button <Detail> to open the
detailed setting screen.
When the Comment <Comment> is [M3] button <Edit> to open the comment input screen.
Содержание AL120-6Series
Страница 2: ...This page intentionally left blank ...
Страница 9: ...Operation Maintenance Manual Introduction Safety Precautions i 7 Page AL120 86 Series Unit mm ...
Страница 16: ...Introduction Conformity Standards Operation Maintenance Manual i 14 Page ...
Страница 17: ...Operation Maintenance Manual Introduction Conformity Standards i 15 Page ...
Страница 18: ...Introduction Conformity Standards Operation Maintenance Manual i 16 Page This page intentionally left blank ...
Страница 20: ...Maintenance Manual Table of Contents C 2 This page intentionally left blank Page ...
Страница 56: ...Test Programs Maintenance Manual 3 26 Page This page intentionally left blank ...
Страница 64: ...Replacement of Consumables Maintenance Manual 4 8 Page This page intentionally left blank ...
Страница 65: ...This page intentionally left blank ...