Test Programs
Maintenance Manual
3-14
Page
Test No./
Test Name
Function Outline
•
[M2]: Moves to the retracted position in front of cassette (Standby position) <Standby>
The A-arm moves vertically up to the cassette height.
The A-arm moves horizontally to the retracted position before cassette.
•
[M3]: Cassette position <Elevator>
The A-arm moves vertically up to the cassette height.
The A-arm moves horizontally to the cassette position.
4-way button [RIGHT]: Move toward the elevator by one pulse
4-way button [LEFT]: Move toward the center table by one pulse
[M1]: Move toward the elevator by 5 mm / Move toward the stage by 5 mm
[M2]: Turn excitation OFF
The state of the sensor is displayed where the wafer No. is shown.
[No. 1]: Origin sensor detection ON:OFF
If a cassette is set on the elevator, operation does not start.
Press the [Quit] button, and the loader is initialized and stops the operation.
[M2]: A-arm vertical direction<A UP/DOWN>
Select the position to check from the submenu.
[M1]:Origin position<Origin>
[M2]: Cassette height position <Cassette position>
[M3]: Stage height position <Stage position>
The sensor and motor origin signal states are indicated by the LEDs on the Wafer No. selector buttons.
[No. 1]: Origin sensor detection ON:OFF
[No. 2]: Motor driver origin ON:OFF
Press the [Quit] button, and the loader is initialized and stops the operation.
---------------------------------------------------------------------------------------------------------------------------------
•
[M1]: Origin position <Origin>
The sensor detection state can be checked by moving the A-arm vertically.
[M1]: Initialize
[M2]: Move up by 3.5 mm
[M3]: Move down by 1.5 mm
[M4]: Switch between excitation ON and OFF
4-way button [UP]: Move upward by one pulse
4-way button [DOWN]: Move downward by one pulse
•
[M2]: Cassette insertion height position <Cassette>
The sensor detection state can be checked by moving the A-arm vertically.
[M1]: Alternately move to the cassette insertion height and a position 0.5 mm lower
4-way button [UP]: Move upward by 55
um
4-way button [DOWN]: Move downward by 55
um
[No.1]: Cassette height sensor detection ON:OFF
•
[M3]: Stage height position <Stage position>
The sensor detection state can be checked by moving the A-arm vertically.
4-way button [UP]: Move upward by one pulse
4-way button [DOWN]: Move downward by one pulse
[No.1]: Cassette height sensor detection ON:OFF
Содержание AL120-6Series
Страница 2: ...This page intentionally left blank ...
Страница 9: ...Operation Maintenance Manual Introduction Safety Precautions i 7 Page AL120 86 Series Unit mm ...
Страница 16: ...Introduction Conformity Standards Operation Maintenance Manual i 14 Page ...
Страница 17: ...Operation Maintenance Manual Introduction Conformity Standards i 15 Page ...
Страница 18: ...Introduction Conformity Standards Operation Maintenance Manual i 16 Page This page intentionally left blank ...
Страница 20: ...Maintenance Manual Table of Contents C 2 This page intentionally left blank Page ...
Страница 56: ...Test Programs Maintenance Manual 3 26 Page This page intentionally left blank ...
Страница 64: ...Replacement of Consumables Maintenance Manual 4 8 Page This page intentionally left blank ...
Страница 65: ...This page intentionally left blank ...