Test Programs
Maintenance Manual
3-18
Page
Test No./
Test Name
Function Outline
•
The sensor state is indicated by the Wafer No. selector button.
The LED on the Wafer No. selector button lights up on the detection of the sensor.
[No. 1]: Origin sensor detection
•
Press the [Quit] button,
and the L-arm returns to its initial position and the loader stops the
operation.
No. 17
Stage
transfer
position
adjustment
<Position adjustment to the stage>
This program can be used to check the position for transferring wafers onto the stage.
Press the [Start] button.
Using the 4-way button, select the number of wafer type to adjust or register from the list of wafer type
numbers displayed on the liquid crystal panel, and then press the [Start] button.
The wafer transfer method can be corrected and adjusted by using the menu buttons on the liquid
crystal panel.
[M1]: Change the A-arm <A-arm change>
[M2]: Transfer a wafer onto the stage <wafer Load>
[M3]: Register the correction value <Save>
Correction direction
4-way button [UP]
: Far side <Far>
4-way button [DOWN]
: Near side <Near>
4-way button [LEFT] : Left <LEFT>
4-way button [RIGHT]
: Right <RIGHT>
Press the [M2] button and then the [Start] button to start the process of transferring a wafer to the
stage.
Before transferring a wafer to the stage, set the orientation flat on the left side.
Rotate the wafer on the stage by using the stage table rotation knob, and check the shift direction.
Input the correction direction by using the 4-way button.
Press the [M3] button to register the correction value.
The registered correction data is saved in the internal memory, and the wafer is unloaded into the
cassette.
Press the [M2] button and then the [Start] button to start the process of unloading the wafer into the
cassette.
Repeat steps
and
to check the decentering amount.
Press the [Quit] button to unload the wafer into the cassette.
Press the [M1] button and then the [Start] button to replace one A-arm with the other A-arm.
Perform steps
to
.
•
Press the [Quit] button to terminate the adjustment.
Содержание AL120-6Series
Страница 2: ...This page intentionally left blank ...
Страница 9: ...Operation Maintenance Manual Introduction Safety Precautions i 7 Page AL120 86 Series Unit mm ...
Страница 16: ...Introduction Conformity Standards Operation Maintenance Manual i 14 Page ...
Страница 17: ...Operation Maintenance Manual Introduction Conformity Standards i 15 Page ...
Страница 18: ...Introduction Conformity Standards Operation Maintenance Manual i 16 Page This page intentionally left blank ...
Страница 20: ...Maintenance Manual Table of Contents C 2 This page intentionally left blank Page ...
Страница 56: ...Test Programs Maintenance Manual 3 26 Page This page intentionally left blank ...
Страница 64: ...Replacement of Consumables Maintenance Manual 4 8 Page This page intentionally left blank ...
Страница 65: ...This page intentionally left blank ...