A-6
Appendix
A-6
6600-0239-000 100
intermittent loading
Duty cycle:
1 minute on,
10 minutes off
Protection against hazards of explosion:
Not protected
Protection against ingress of liquids:
Not protected
Degree of electrical connection
Equipment without
between equipment and patient:
an applied part
Degree of mobility:
Mobile
Electromagnetic
Meet requirements of IEC 601-1-2
Compatibility
ThermaLink Option Specifications
(Available on controllers with serial numbers beginning with HBJ and HCE)
A. Hewlett Packard CMS Monitor (Vuelink)
The incubator may be connected to a Hewlett Packard CMS Monitor via a Vuelink open interface
module (HP P/N M1032A Option A05) and HP Ohmeda option cable. The
parameters passed to the HP monitor are: patient temperature, air temperature, set
temperature, incubator power, and D.E.T. (Desired Environmental Temperature); the first three
parameters are displayed by default on power-up. The user can select to display any or all of
the 5 parameters. All alarms are also passed to the monitor. See the Hewlett Packard CMS and
Vuelink manuals for features and functions of the HP system.
B. SpaceLabs Monitor (Flexport)
The Incubator may be connected to a SpaceLabs Bedside Monitor via a Flexport
(SL P/N 90444A). The parameters passed to the SpaceLabs Bedside Monitor are: patient
temperature, air temperature, set temperature, and P-A (Patient minus Air). All alarms
are also passed to the monitor. See the SpaceLabs Monitor and Flexport manuals for features
and functions of the SL system
C. RS-232 Serial data
WARNING
The computer or RS-232 monitor’s user program must continuously check the data link.
The program should constantly verify connection to the incubator controller and check for
updated data.
Note: In the event of a power failure, all serial communication will cease until power
is restored.
RS-232 Connector
The Nurse Call and the serial data output share the same female, nine pin, d-type
connector.
Pin 2: Receive Data (incubator input)
Pin 3: Transmit Data (incubator output)
Pin 5: Gnd (Signal Ground)
Cable requirements
The user interface cable must have capacitance less than 1500 pF. It should be a
shielded cable such as Belden 9611 with AMP shielding kit 748046-1 and ferrule 747579-8.
Data transmission
The incubator continuously sends data from the time that it is first powered up.
Note that the incubator cannot be controlled through the serial port. Data output stops when
the incubator receives a <cntrl>S (XOFF) and resumes when it receives a <cntrl>Q (XON).
Содержание Ohio Care Plus
Страница 1: ...Ohio CarePlus Incubator and Ohio CarePlus Access Incubator Operation and Maintenance Manual...
Страница 6: ...Table of Contents iv iv 6600 0239 000 100 Notes...
Страница 10: ...1 1 1 Introduction 1 4 6600 0239 000 100 Notes...
Страница 16: ...2 6 2 Getting Started 2 6 6600 0239 000 100 Notes...
Страница 26: ...1 4 1 Introduction 1 4 6600 0239 000 100 Notes...
Страница 54: ...5 18 5 18 6600 0239 000 100 Notes...
Страница 66: ...7 4 7 Service Procedures 7 4 6600 0239 000 100 Notes...
Страница 70: ...8 Illustrated Parts 8 4 8 4 6600 0239 000 100 Notes...
Страница 80: ...A 10 Appendix A 10 6600 0239 000 100...