3-6
3/General Information
3-6
6600-0239-000 100
The Override switch lets you select air control temperatures higher
than 37.0°C (98.6°F). To raise the air control temperature above
37°C, increase the control temperature to 37°C, press the Enable
switch, press the Override switch, and then increase the air control
temperature. The maximum set temperature is 39°C. The override
indicator remains illuminated to indicate that the air control temperature
exceeds 37.0°C.
Pressing the Patient Control switch selects the Patient control mode of
operation and illuminates the Patient Control indicator (next to Control
Temperature display). To prevent accidental mode changes, you must
press the Enable switch to activate the Patient Control switch. The
audible alarm sounds briefly when the mode of operation is changed.
The first time that you select the patient control mode, the control
temperature display also flashes and an operator prompt tone sounds
until you enter a control temperature.
These switches adjust the control temperature. To prevent accidental
changes, you must press the Enable switch to activate these switches.
These switches remain active as long as the enable indicator is
illuminated, approximately 12 seconds after the last switch is pressed.
For controllers with serial numbers beginning with HBJ or HCE, the power
switch is located on the right side of the controller assembly, adjacent to
the power cord socket. It switches the incubator power On and Off. Two
circuit breakers, located within the controller, limit the maximum current
drawn by the incubator. If the circuit breaker trips, remove the unit from
use and contact qualified service personnel for repair.
On all other controllers, combination power switch/circuit breaker is
located on the right side of the controller assembly, adjacent to the
power cord socket. It switches incubator power On and Off. It also
resets the circuit breaker, which limits the maximum current drawn by
the incubator. If the circuit breaker trips, the power switch returns to the
off (O) position. To reset the circuit breaker, set the switch back to the
on (|) position. If the circuit breaker trips again, remove the unit from
use and contact qualified service personnel for repair.
Alarms
Note: On Controllers with serial numbers beginning with HBJ or HCE, all silenceable
alarms are preceded by a 30 second operator prompt tone.
Seven LEDs on the left side of the control panel illuminate for various alarm conditions.
Each visual indicator is accompanied by an audio signal.
Air
Control
Air
Override
> 37
°
C
Patient
Control
Figure 3-3
Alarm indicators
Alarms
Patient Temp
Control Temp
High Air Temp
Air Circulation
Probe Failure
System Failure
Power Failure
CI.02.215
Содержание Ohio Care Plus
Страница 1: ...Ohio CarePlus Incubator and Ohio CarePlus Access Incubator Operation and Maintenance Manual...
Страница 6: ...Table of Contents iv iv 6600 0239 000 100 Notes...
Страница 10: ...1 1 1 Introduction 1 4 6600 0239 000 100 Notes...
Страница 16: ...2 6 2 Getting Started 2 6 6600 0239 000 100 Notes...
Страница 26: ...1 4 1 Introduction 1 4 6600 0239 000 100 Notes...
Страница 54: ...5 18 5 18 6600 0239 000 100 Notes...
Страница 66: ...7 4 7 Service Procedures 7 4 6600 0239 000 100 Notes...
Страница 70: ...8 Illustrated Parts 8 4 8 4 6600 0239 000 100 Notes...
Страница 80: ...A 10 Appendix A 10 6600 0239 000 100...