6-8
6/Maintaining the Incubator
6-8
6600-0239-000 100
Porthole
The following cleaning agents are acceptable for cleaning and disinfecting the porthole:
Isopropyl Alcohol, 70% concentration
Hydrogen Peroxide, 6% concentration (such as Endo-Spor 100% strength)
Cavicide Spray, 100% concentration as sprayed from the can (active ingredients -
Diisobutylphenoxyethyl dimethyl benzyl ammonium chloride .25%, Isopropanol 14.85%).
Idophor Solution, 0.27% concentration (active ingredients - Polythoxy polypropoxy
polyethoxy ethanoliodine complex 9.10%, nonyl phenoxypoly [ethyleneoxy]
ethanoliodine 8.74%, provides 1.6% minimum Titratable iodine, such as
Wescodyne- mixed 2 fl. oz. to 1 gal. H
2
O) Note: this solution will stain porthole
seals and bumpers yellow.
The following cleaning agents will damage the material being cleaned and are not
recommended:
Sodium Hypochlorite (bleach)
Quaternary Ammonium (such as Virex)
Glutaraldehyde (such as Cidex Plus)
Compartment Temperature Probe
Wipe the temperature probe with a cloth dampended with a mild detergent-disinfectant
solution, such as Cavicide™. Be sure to allow the probe to thoroughly dry before
putting the incubator back into operation.
Other Clear Plastic Parts
CAUTIONS
To minimize the generation of static electricity, do not polish the incubator hood
with a dry cloth.
Do not autoclave or gas sterilize the mattress or any of the plastic parts.
Do not use strong acids or strong bases on plastic parts.
Clean the hood, the inner wall, and other clear plastic parts of the incubator with a
disinfectant-detergent. Make sure that the inner walls have been removed so that all
incubator surfaces are exposed.
Apply the cleaning solution with a clean cloth or sponge. Dry the plastic parts with a
clean damp soft cloth to avoid scratches.
Note: Do not allow excess cleaning solution to seep in between plastic parts (for
example: between the hood and porthole, or between the hood and the inner wall
retainers) where it can not be easily wiped dry with a cloth.
Lower unit (mattress, mattress tray, base platform cover, etc.)
Clean the base platform cover, the mattress and the mattress tray with a mild
detergent-disinfectant solution. The tilt mechanism parts in the patient compartment
can be cleaned by immersing them in soap and water or a mild detergent disinfectant
solution. Clean the exterior tilt controls by wiping down the exposed surfaces with a
cloth dampened with a mild detergent-disinfectant solution. Do not immerse exterior
tilt mechanism. Rinse and thoroughly dry the parts before reassembly.
Oxygen inlet
Clean the oxygen inlet with a mild cleaning solution or disinfectant.
Cabinet
Clean the cabinet with a mild cleaning solution. Apply the cleaning solution with a clean
cloth or sponge. Dry the cabinet with a clean, soft cloth.
Содержание Ohio Care Plus
Страница 1: ...Ohio CarePlus Incubator and Ohio CarePlus Access Incubator Operation and Maintenance Manual...
Страница 6: ...Table of Contents iv iv 6600 0239 000 100 Notes...
Страница 10: ...1 1 1 Introduction 1 4 6600 0239 000 100 Notes...
Страница 16: ...2 6 2 Getting Started 2 6 6600 0239 000 100 Notes...
Страница 26: ...1 4 1 Introduction 1 4 6600 0239 000 100 Notes...
Страница 54: ...5 18 5 18 6600 0239 000 100 Notes...
Страница 66: ...7 4 7 Service Procedures 7 4 6600 0239 000 100 Notes...
Страница 70: ...8 Illustrated Parts 8 4 8 4 6600 0239 000 100 Notes...
Страница 80: ...A 10 Appendix A 10 6600 0239 000 100...