6-1
6/Maintaining the Incubator
6-1
6600-0239-000 100
6/Maintaining the Incubator
WARNINGS
The heater is hot enough to cause skin burns. Unplug the incubator and
allow the heater to cool before disassembly or cleaning.
Always disconnect the power before performing service or maintenance
procedures detailed in this manual. Apply power only if you are specifically
instructed to do so as part of the procedure.
Thoroughly air dry the incubator after cleaning it with flammable agents.
Small amounts of flammable agents, such as ether, alcohol or similar
cleaning solvents left in the incubator can cause a fire.
Never oil or grease oxygen equipment unless a lubricant that is made and
approved for this type of service is used. Oils and grease oxidize readily,
and in the presence of oxygen, will burn violently.
In this section
Cleaning schedule ..............................................................................................................................................6-1
Disassembling the incubator for a complete cleaning ..........................................................................6-2
Cleaning and disinfecting individual components .................................................................................6-5
Humidifier ...................................................................................................................................................6-6
Patient Probe (Reusable) .......................................................................................................................6-6
Controller assembly ................................................................................................................................6-7
Compartment Temperature Probe ....................................................................................................6-8
Porthole .......................................................................................................................................................6-8
Other Clear Plastic Parts ........................................................................................................................6-8
Lower unit (mattress, mattress tray, base platform cover, etc.) ...............................................6-8
Oxygen inlet ..............................................................................................................................................6-8
Cabinet ........................................................................................................................................................6-8
Cleaning schedule
This schedule lists the minimum frequencies. Always follow hospital and local
regulations for required frequencies.
Weekly or after each patient
Disinfect the humidifier if it was used.
Perform a complete cleaning and if necessary disinfect the entire incubator.
Check the air filter; replace the air filter after use with an infectious patient.
Check the front door and the porthole latches and hinges to make sure that they will
not open accidently.
Quarterly
Replace the air filter.
When the incubator is not in use, use standard analytical methods to make sure that
the infant compartment and any direct contact components, such as the humidifier
and the oxygen inlet, meet hospital cleanliness standards.
Содержание Ohio Care Plus
Страница 1: ...Ohio CarePlus Incubator and Ohio CarePlus Access Incubator Operation and Maintenance Manual...
Страница 6: ...Table of Contents iv iv 6600 0239 000 100 Notes...
Страница 10: ...1 1 1 Introduction 1 4 6600 0239 000 100 Notes...
Страница 16: ...2 6 2 Getting Started 2 6 6600 0239 000 100 Notes...
Страница 26: ...1 4 1 Introduction 1 4 6600 0239 000 100 Notes...
Страница 54: ...5 18 5 18 6600 0239 000 100 Notes...
Страница 66: ...7 4 7 Service Procedures 7 4 6600 0239 000 100 Notes...
Страница 70: ...8 Illustrated Parts 8 4 8 4 6600 0239 000 100 Notes...
Страница 80: ...A 10 Appendix A 10 6600 0239 000 100...