4-8
4/Preoperative Checkout Procedure
4-8
6600-0239-000 100
Figure 4-1
Power connections and control locations
Operating the Elevating Base
Note: Continuously depressing the
M
or
?
pedals for long periods can cause the
motor to overheat. To avoid damage, a thermal switch opens cutting the motor
power. The switch remains open, preventing operation, until the motor cools.
1. Complete the Checkout procedure .
2. Use the
M
and
?
pedals to adjust the Elevating Base to the desired height.
WARNING
When a patient occupies the incubator, always close the incubator door and
portholes before raising or lowering the Elevating Base.