6-2
6/Maintaining the Incubator
6-2
6600-0239-000 100
Figure 6-1
Removing the front door
Disassembling the incubator for a complete cleaning
If the incubator has been exposed to body fluids, check your hospital’s infection
control procedures. You may need to wear protective clothing and eye wear or use
special sterilization agents and cleaning procedures.
WARNING
Disconnect the power cord before disassembling or reassembling the
lower unit.
CAUTIONS
Electronic devices in the microprocessor controller are susceptible to damage
from discharges of static electricity. These devices are adequately protected,
but can be damaged if the unit is disassembled beyond that recommended for
cleaning and maintenance.
Do not remove the air temperature sensor mounted to the rear hood wall; the
air temperature sensor must remain in the correct position for proper incubator
operation.
1. Turn the power switch Off and unplug the unit from the power outlet.
2. If the incubator was previously on, allow it to cool for at least 30 minutes.
3. Remove the filter by unscrewing the two filter mounting knobs on either side of
the vented rear panel. Pull off the panel and remove the filter.
WARNING
The air filter cannot be cleaned and must be replaced every three months,
when visibly dirty or after use with an infectious patient. If the air filter is
exposed to liquids or disinfectants, replace it with a new filter.
4. Unplug the power cord from the right side of the controller.
5. Unplug the air temperature sensor and the patient probe from the controller.
6. Pull up on both controller latches and slide the controller forward, out of the lower
unit. Be careful not to bump the heater or the fan. They may grate against each
other or the base platform if they are knocked out of alignment.
7. Open the front door to reveal the two spring loaded hinge pins which attach the
door to the lower unit. Pull both pins out toward the sides of the hood and lift off
the door. See Figure 6-1.
Hinge Pins
CI.02.015
Содержание Ohio Care Plus
Страница 1: ...Ohio CarePlus Incubator and Ohio CarePlus Access Incubator Operation and Maintenance Manual...
Страница 6: ...Table of Contents iv iv 6600 0239 000 100 Notes...
Страница 10: ...1 1 1 Introduction 1 4 6600 0239 000 100 Notes...
Страница 16: ...2 6 2 Getting Started 2 6 6600 0239 000 100 Notes...
Страница 26: ...1 4 1 Introduction 1 4 6600 0239 000 100 Notes...
Страница 54: ...5 18 5 18 6600 0239 000 100 Notes...
Страница 66: ...7 4 7 Service Procedures 7 4 6600 0239 000 100 Notes...
Страница 70: ...8 Illustrated Parts 8 4 8 4 6600 0239 000 100 Notes...
Страница 80: ...A 10 Appendix A 10 6600 0239 000 100...