A-2
Appendix
A-2
6600-0239-000 100
Performance specifications
Patient temperature measurement
Range Displayed: +22 to 42°C (71.6 to 107.6°F)
Accuracy*: ± 0.3°C (± 0.5°F) within a range of 30-42°C
Resolution: ± 0.1°C or °F
Probe Model Numbers: LA003, LA005 (disposable)
Probe Interchangeability: ± 0.1°C (± 0.2°F)
Air temperature measurement
Range Displayed: 5 to 50°C (41.0 to 122.0°F)
Resolution: ± 0.1°C or °F
Accuracy*: Varies over temperature range
Temperature Range
Accuracy
5 to 22°C (41-71.6°F)
± 0.5°C (0.9°F)
22.0 to 42.0°C (71.6-107.6°F)
± 0.3°C (0.5°F)
42.0 to 50.0°C (107.6-122.0°F)
± 0.5°C (0.9°F)
Control temperature ranges
Patient Control Mode: 35.0 to 37.0°C (95.0 to 98.6°F), up to 37.5°C (99.5°F) with
internal adjustment
Air Control Mode: 20.0 to 37.0°C (68.0 to 98.6°F), up to 39.0°C (102.2°F) with
control panel Override switch.
Thermal performance
Temperature Rise Time*: Less than or equal to 20 min
Temperature Variability*: 0.2°C (0.4°F)
Temperature Overshoot*: Less than or equal to 0.8°C (1.4°F)
Note: * These terms are defined in the Glossary.
Alarms
Indicator and Nonsilenceable Audio Alarm
Power Failure
Indicator and Silenceable Single Tone Alarm
Difference between patient temperature and patient control temperature exceeds
1.0°C patient control mode (can be adjusted to 0.5°C).
Air temperature is 1.5°C above or 3.0°C below air control temperature (air control mode;
Disabled for 15 minutes after control temperature change and for 30 minutes
on power up).
Indicator, Silenceable Two Tone Alarm and Heater Shutdown
Air temperature sensor disconnected (both modes)
Patient probe disconnected or malfunctioning (patient control mode)
Air circulation system failure
Patient temperature > 42.0°C or < 30.0°C
Air temperature > 38.0°C (air control mode w/o Override)
Air temperature > 40.0°C (patient control mode or air control mode with Override)
Содержание Ohio Care Plus
Страница 1: ...Ohio CarePlus Incubator and Ohio CarePlus Access Incubator Operation and Maintenance Manual...
Страница 6: ...Table of Contents iv iv 6600 0239 000 100 Notes...
Страница 10: ...1 1 1 Introduction 1 4 6600 0239 000 100 Notes...
Страница 16: ...2 6 2 Getting Started 2 6 6600 0239 000 100 Notes...
Страница 26: ...1 4 1 Introduction 1 4 6600 0239 000 100 Notes...
Страница 54: ...5 18 5 18 6600 0239 000 100 Notes...
Страница 66: ...7 4 7 Service Procedures 7 4 6600 0239 000 100 Notes...
Страница 70: ...8 Illustrated Parts 8 4 8 4 6600 0239 000 100 Notes...
Страница 80: ...A 10 Appendix A 10 6600 0239 000 100...