2-3
2/Getting Started
2-3
6600-0239-000 100
Mounting the Care Plus on the cabinet
WARNING:
Safely mounting or dismounting the Care Plus Incubator or the Elevating
Base rail system requires two people.
1. Verify that O-rings or foam gaskets are installed in the depression around each of
the incubator mounting holes in the cabinet top.
2. Lift the Care Plus Incubator onto the cabinet. Make sure that the cabinet doors
face the same direction as the incubator’s front door. Refer to Figure 2-2.
3. Install the four mounting knobs.
WARNING
If mounting hardware is not securely fastened, the incubator could tip off the
stand.
CAUTION
When handling the controller, avoid bumping the fan or the heater. If these items
are knocked out of alignment, the fan can grate against the heater or the base.
4. Install the controller. Lift the levers on the controller and slide it into the incubator
as shown in Figure 2-2. Push the levers down to lock the controller in place.
5. Refer to section 4 in this manual to perform preoperative checkout procedures.
Figure 2-2
Mounting the Care Plus Incubator
Mounting
Knob
CI.02.189
Controller
Controller
Lever
Содержание Ohio Care Plus
Страница 1: ...Ohio CarePlus Incubator and Ohio CarePlus Access Incubator Operation and Maintenance Manual...
Страница 6: ...Table of Contents iv iv 6600 0239 000 100 Notes...
Страница 10: ...1 1 1 Introduction 1 4 6600 0239 000 100 Notes...
Страница 16: ...2 6 2 Getting Started 2 6 6600 0239 000 100 Notes...
Страница 26: ...1 4 1 Introduction 1 4 6600 0239 000 100 Notes...
Страница 54: ...5 18 5 18 6600 0239 000 100 Notes...
Страница 66: ...7 4 7 Service Procedures 7 4 6600 0239 000 100 Notes...
Страница 70: ...8 Illustrated Parts 8 4 8 4 6600 0239 000 100 Notes...
Страница 80: ...A 10 Appendix A 10 6600 0239 000 100...