Calibration
W-4510 Weigh Scale Technical Reference Manual
83
Initial calibration overview
Before you can use the gauge, a relationship between the detector signal and the actual
process values must be determined. The initial calibration determines this relationship by
collecting data and measuring the actual process value. The initial calibration of the gauge
includes following steps.
1.
Verifying the correct measurement units
2.
Verifying the proper measurement span entries
3.
Determining the linearization
4.
Collecting data for the zero and span
5.
Calculating the calibration
6.
Resetting the absorber value
7.
Saving the calibration information to EEPROM
Figure 39: Graphical representation of the calibration curve
Counts
Cal_lo
Cal_hi
Weight_lo Weight_hi
Weight
Содержание W-4510
Страница 9: ...Preface viii W 4510 Weigh Scale Technical Reference Manual Notes...
Страница 15: ...Preface xiv W 4510 Weigh Scale Technical Reference Manual Notes...
Страница 17: ...Preface xvi W 4510 Weigh Scale Technical Reference Manual Notes...
Страница 29: ...Introduction 12 W 4510 Weigh Scale Technical Reference Manual Notes...
Страница 36: ...Installation W 4510 Weigh Scale Technical Reference Manual 19 Figure 6 Typical Smart Pro dimensions...
Страница 48: ...Installation W 4510 Weigh Scale Technical Reference Manual 31 Figure 13 W 4510 terminal illustration...
Страница 125: ...Calibration 108 W 4510 Weigh Scale Technical Reference Manual Notes...
Страница 145: ...Diagnostics and repair 128 W 4510 Weigh Scale Technical Reference Manual Notes...
Страница 155: ...Appendix I Parameter blocks 138 W 4510 Weigh Scale Technical Reference Manual Notes...