Diagnostics and repair
W-4510 Weigh Scale Technical Reference Manual
117
Figure 58: Pro Pac Analog Input-2
Analog Input-2
Start
Is
signal
present at
Input 2 TB1
(7-8)
?
Check
jumpers, I/O
signal cable
OK?
Have
power supply
levels been
checked?
Replace CPU
End
Refer to power
supply flow chart
Problem with
external wiring or
transmitter
Correct
Yes
No
Yes
Yes
No
No