Setup
48
W-4510 Weigh Scale Technical Reference Manual
Analog output setup
065 ANALOG OUTPUTS
SET ANALOG OUT SPAN
CAL ANALOG OUTPUTS
Figure 23: Analog output setup screens
There are two analog output channels. You must span and calibrate the channels before they
can transmit a meaningful signal. The analog output span can be set independent from the
measurement span of this gauge. Setting the span is simply a matter of defining what
process values that correspond to the 4 and 20mA or 0 and 20mA levels.
The Smart Pro or Pro PAC CPU board mounts on the door of the Smart Pro. The jumpers
are clearly labeled. See Figure 24 for details on jumper and test points.