Setup
76
W-4510 Weigh Scale Technical Reference Manual
Adaptive filter
After the measurement channel comes out of the cutoff condition, the output of the
measurement filter is set to the first reading seen. Because of the normal fluctuations from
measurement noise and the nature of radioactive decay, this first reading may not be
precisely representative of the process conditions. To minimize this effect, you can activate
an adaptive filter that improves the performance of the gauge. This adaptive filter is set up in
screen
529
, addresses
90
through
96
.
Procedure 18: Setting up an adaptive filter
1. From
529 APP CODE ACCESS
screen, address
90,
type
1
and press ENTER
to enable the adaptive filter.
2. From
529 APP CODE ACCESS
screen, address
91
, type the minimum value to
set the time constant when coming of fast cutoff (default minimum value is set
to a value of one) and press ENTER
3. From
529 APP CODE ACCESS
screen, address
92
, type the maximum value
to increment the time constant (set same as normal time constant up to a value
of 100 seconds, the default value is set to 30 seconds) and press ENTER
4. From
529 APP CODE ACCESS
screen, address
93
, type how many seconds
to maintain each time constant before bumping to the new value (default value
is set to 10 seconds) and press ENTER
5 From
529 APP CODE ACCESS
screen, address
94
, Enter how much to step
up the time constant (default value is five seconds) and press ENTER
6.
To view the current time constant value of the adaptive filter, go to
529 APP
CODE ACCESS
screen, address
95
7.
Connect write protect jumper, JP17 and save to EEPROM
8.
Disconnect the write protect jumper, JP17. This protects the information in the
EEPROM.
529 APP CODE ACCESS
screen, address
95
and
96
are diagnostic locations. Address
95
is
the current time constant with the adaptive filter. Address
96
is the maximum time constant
minus the current time constant. When this number gets to zero, the adaptive filtering is
complete until the detection of the next cutoff condition.
Содержание W-4510
Страница 9: ...Preface viii W 4510 Weigh Scale Technical Reference Manual Notes...
Страница 15: ...Preface xiv W 4510 Weigh Scale Technical Reference Manual Notes...
Страница 17: ...Preface xvi W 4510 Weigh Scale Technical Reference Manual Notes...
Страница 29: ...Introduction 12 W 4510 Weigh Scale Technical Reference Manual Notes...
Страница 36: ...Installation W 4510 Weigh Scale Technical Reference Manual 19 Figure 6 Typical Smart Pro dimensions...
Страница 48: ...Installation W 4510 Weigh Scale Technical Reference Manual 31 Figure 13 W 4510 terminal illustration...
Страница 125: ...Calibration 108 W 4510 Weigh Scale Technical Reference Manual Notes...
Страница 145: ...Diagnostics and repair 128 W 4510 Weigh Scale Technical Reference Manual Notes...
Страница 155: ...Appendix I Parameter blocks 138 W 4510 Weigh Scale Technical Reference Manual Notes...