Calibration
W-4510 Weigh Scale Technical Reference Manual
85
Step 3: Determine the linearization
Nuclear Measurements are inherently non-linear. The heavier the loading, the more the
measurement becomes non-linear. Since the gauge calibrates near the ends of the
measurement span, the gauge reads correctly there. However, you must load the correct
linearization curve into the Smart Pro before the gauge reads the correct weight in the middle
of the span. This curve depends on the following components:
•
Measurement span
•
Geometry of the source and detector
Procedure 24: Verify the curve is loaded into Smart Pro
1. From
the
052 ENTER PASSWORD
screen, enter the password for the appropriate
access level
2.
Go to screen
528 CURVE #
, to view the current values from 1-40
3.
Enter the item number and press ENTER to view the specific data point.
4.
Check items 0 to 40. If the data is between 0 and 10,000 (0–100%) in increments
of 250, then Linearization is not present
5.
Several loadings between 0–100% must be run on the belt or simulated in a static
condition to derive data for this curve. Upon completing the data collection, provide
Ohmart with the collected data points. Ohmart calculates the curve and supplies
41 linearization points.
Содержание W-4510
Страница 9: ...Preface viii W 4510 Weigh Scale Technical Reference Manual Notes...
Страница 15: ...Preface xiv W 4510 Weigh Scale Technical Reference Manual Notes...
Страница 17: ...Preface xvi W 4510 Weigh Scale Technical Reference Manual Notes...
Страница 29: ...Introduction 12 W 4510 Weigh Scale Technical Reference Manual Notes...
Страница 36: ...Installation W 4510 Weigh Scale Technical Reference Manual 19 Figure 6 Typical Smart Pro dimensions...
Страница 48: ...Installation W 4510 Weigh Scale Technical Reference Manual 31 Figure 13 W 4510 terminal illustration...
Страница 125: ...Calibration 108 W 4510 Weigh Scale Technical Reference Manual Notes...
Страница 145: ...Diagnostics and repair 128 W 4510 Weigh Scale Technical Reference Manual Notes...
Страница 155: ...Appendix I Parameter blocks 138 W 4510 Weigh Scale Technical Reference Manual Notes...