OFITE, 11302 Steeplecrest Dr., Houston, TX 77065 USA / Tel: 832-320-7300 / Fax: 713-880-9886 / www.ofite.com
33
UCA
Cell Assembly
The cell body and both cell caps were manufactured and pressure tested
together. All three pieces are serialized. Before assembling the test cell, make
sure all three pieces have the same serial number.
For a complete diagram of the test cell, refer to page 74.
Cell Cap Assembly
1. Apply a thin layer of high-temperature grease to the inside surfaces of the
two cell caps. Wipe off any excess grease.
2. Apply high-temperature grease to both cell cap o-rings (#123-011).
3. Place an o-ring in the grooves of each cell cap.
4. Make sure the transducers and the transducer holes in the cell caps are
clean and free of debris. They can be cleaned with a rag or paper towel.
Alcohol can also be used if further cleaning is necessary.
O-ring Groove
Top UCA Cell Cap
Bottom Cell Cap