Calibration
SGSM Load Cell
The load cell on the SGSM should be calibrated periodically to ensure
accurate readings. The calibration is performed with a dead-weight set.
1. Remove the cover of the SGSM Head Assembly.
2. Remove the torque spring and install the calibration pulley.
3. Connect the wire to the torque spring post and thread it over the
calibration pulley.
Torque Spring Post
Calibration Pulley
Load Cell
Torque Spring