68
Calibrating the Thermocouple requires the use of a calibrated heating device.
1. Start the software.
2. Connect the thermocouple to a dry block calibrating device.
3.
Set the Pressure switch to Off.
4. Access the thermocouple calibration:
a.
In the software select: Utilities → Cell 1 or Cell 2 → Calibrate →
Thermocouple
b. In the Onboard Display, go to the Calibrate screen and choose
“Temperature”.
5. Follow the instructions on the screen. You will be prompted to set the
temperature on the dry block to various set points. When the temperature
stabilizes, click the “Accept” (software) or “>” (onboard) button.
6. When all steps are complete, you will see the calibration data, including a
Pass or Fail.
7. Click the “Save” (software) or “Apply” (onboard) button.
Calibration
Thermocouple