Maintenance
Fuses
The instrument has two fuses. If the unit will not power on, check both fuses
to make sure they are not blown.
1. Remove the back panel and inspect the fuses.
2. If a fuse is blown, a red LED light will come on below the housing.
3. Unplug the power cord leading to the UCA.
4. Depress the tab on the bottom of the fuse housing and flip the housing up.
5. Open the door on the side of the fuse housing.
6. Replace the blown fuse.
7. Close the door on the fuse housing and push the housing down back into
place.
8. Plug the power cord back into the UCA.
9. Check the system to ensure it is working again.
Fuse Housings
LED Lights