OFITE, 11302 Steeplecrest Dr., Houston, TX 77065 USA / Tel: 832-320-7300 / Fax: 713-880-9886 / www.ofite.com
8
Setup
1.
Carefully remove the instrument from the crate and place it on a flat,
stable surface. Make sure to allow for adequate air flow around the unit,
especially near the vents on the sides and back.
To ensure personal safety, always use proper lifting techniques.
Position the unit such that the user can quickly disconnect all electrical
plugs in case of an emergency.
2.
Make sure all switches are off and the valves on the front are turned to
DRAIN.
3. Connect the
AIR, CHILL IN, CHILL OUT, WATER SUPPLY,
and
DRAIN
connectors on the back of the unit to their appropriate source and plug in
the power cords.
Make sure the “Drain” and “Chill Out” are suitable for high temperature
discharge.
Make sure the compressed air supply is clean and dry.
4. Connect the UCA to the PC. The UCA can connect to the PC via serial
(RS-232), Ethernet, or USB.
Do not connect an Internet device (PC, laptop, wall-mounted Ethernet
port, etc.) to the SGSM port. This could damage your devices.
5. If you will be running an SGSM test, connect the SGSM head to the
corresponding port on the back of the UCA with the supplied Ethernet
cable. The cell on the left is SGSM 1 and the cell on the right is SGSM 2.
6. Make sure the HEAT, COOL, and PRESSURE switches are all in the OFF
position before turning on the main power.
7. Turn the “Main Power” switch on.
8. Open the software and connect using the device manager (page 11).
Ensure that the live data on the front panel displays matches the data
displayed on the software.
Ethernet
Cable