OFITE, 11302 Steeplecrest Dr., Houston, TX 77065 USA / Tel: 832-320-7300 / Fax: 713-880-9886 / www.ofite.com
43
9. Install the two bushings. One bushing goes into the hole in the top of the
top cell cap. The other goes inside the Mounting Adapter.
Make sure the bushings sit flush with the metal.
10. Secure the mounting adapter (120-58-04) with the short end facing up.
11. Place the upper backup ring (#120-53-58) around the top of the mounting
adapter with the o-ring (#120-53-32) on top of it. The backup ring should
be oriented with the taper pointing away from the o-ring.
12. Place a retaining ring (#120-53-82) on the mounting adapter to hold the
o-ring in place.
13. Secure the mounting adapter with the longer end of the
shaft pointed up. Place the lower backup ring (#120-53-
59) around the bottom of the mounting adapter with the
o-ring (#120-53-31) on top of it. The backup ring should be
oriented with the taper pointing away from the o-ring.
14. Install the retainer ring (#120-58-05) above the o-ring.
Bushing
(#120-53-42)
Bushing
(#120-53-42)
Backup Ring
(#120-53-58)
O-ring
(#120-53-52)
Mounting
Adapter
Retaining Ring (#120-53-82)
High Pressure SGSM Only
Retainer Ring
(#120-58-05)
Mounting Adapter
(#120-58-04)
O-ring
(#120-53-31)
Lower Backup Ring
(#120-53-59)
O-ring and Backup Ring in
Proper Alignment
Metal Seal Ring
(Cross Section)