6
Checking and Adjustment
170
Pulse Laser Station NPL-302 Series Instruction Manual
Checking and Adjusting the Plate Level
6.1
The axis of the plate level vial must be at right angles to the vertical axis of the
instrument.
To check and adjust the plate level:
1.
Set up the instrument on the tripod.
2.
Follow the leveling procedures described in Leveling, page 16.
3.
Rotate the alidade 180°.
4.
Check whether the bubble is in the center of the vial.
5.
If the bubble is not in the center of the vial, adjust the plate level:
a.
Using the adjusting pin supplied, rotate
the plate level’s adjustment screw until
the bubble has moved half of the
distance back to the center.
b.
Using leveling screw A, move the
bubble into the center of the vial.
c.
Checking and Adjusting the Circular Level
6.1
Once you have checked and adjusted the plate
level, check the circular level.
If the bubble is not in the center of the level,
use the adjusting pin to rotate the three
adjustment screws until the bubble is centered.
C
B
A
2
Содержание NPL-302 Series
Страница 1: ...H163E 05 10 TF 9 Pulse Laser Station NPL 302 Series Instruction Manual NPL 332 NPL 352 NPL 362 ...
Страница 4: ......
Страница 5: ...Safety 1 In this chapter Q Introduction Q Laser Safety Q Warnings and Cautions ...
Страница 12: ...Safety xii Pulse Laser Station NPL 302 Series Instruction Manual ...
Страница 28: ...1 Introduction 8 Pulse Laser Station NPL 302 Series Instruction Manual ...
Страница 46: ...2 Preparation 26 Pulse Laser Station NPL 302 Series Instruction Manual ...
Страница 82: ...3 Getting Started 62 Pulse Laser Station NPL 302 Series Instruction Manual ...
Страница 188: ...5 Menu Key 168 Pulse Laser Station NPL 302 Series Instruction Manual ...
Страница 196: ...6 Checking and Adjustment 176 Pulse Laser Station NPL 302 Series Instruction Manual ...
Страница 208: ...8 System Diagrams 188 Pulse Laser Station NPL 302 Series Instruction Manual ...